PIECAL 525B
The PIECAL 525B from Practical Instrument Electronics (PIE) , which replaces the PIE Model 525. This moves their Dual Thermocouple & RTD calibrator to the rugged Evolution style enclosure with a rubber boot and tilt stand for bench use.
​
The PIECAL 525B combines features from the PIECAL 311 Automated RTD Calibrator & PIECAL 422 Automated Thermocouple Calibrator.
​
Other features include:
-
14 Thermocouple types
-
RTD types for Platinum 10, 50, 200 & 500 Ohm for the 1.3850 (IEC 751/DIN) curve
-
More accurate & stable cold junction compensation
NIST Traceable Cold Junction Sensor in isothermal block -
Large, backlit display
Shows millivolts and cold junction temperature while displaying thermocouple temperature.
Shows ohms while displaying RTD temperature -
Troubleshoot RTD Sensors & Instruments
Automatic detection of 2, 3 or 4 wire connections indicates when individual wire connections are open. Verify proper instrument operation by measuring the excitation current
-
Calibrate Thermocouple instruments to 0.1 & 0.01 °F & °C
Types J, K, T, E, R, S, B, N, G, C, D, L (J DIN), U (T DIN) & P (Platinel): -13.000 to 80.000 mV -
Calibrate RTD instruments to 0.1 & 0.01 °F & °C
Pt 10, 50, 100, 200, 500 & 1000 Ohm (3850)
Pt 100 Ohm (alpha=3902, 3916, 3926)
Cu 10 & 50 Ohm and Ni 120 Ohm -
Calibrate with Confidence
Resolution to 0.01 °F & °C; Accurate to 0.2°
Cold Junction Sensor Accurate to 0.05°C -
Perform Heat Treating Surveys & Accuracy Tests
Meets or exceeds the requirement of AMS 2750 as both a Secondary Standard Instrument and as a Field Test Instrument -
Troubleshoot RTD Sensors & Instruments
Automatic detection of 2, 3 or 4 wire connections indicates when individual wire connections are open. Verify proper instrument operation by measuring the excitation current. -
Easily find trip points with
Instantly select three outputs with EZ-CHECK switch
Faster Calibrationstwo speed EZ-DIAL knob & Automatic Stepping -
Easily Change Thermocouples and RTD Types & Scale
Double click to select Thermocouples or RTD type & °F & °C -
Use in the Plant, Field or Shop
Includes Rubber Boot and mV Test Leads
Low profile switches resist accidental damage
Brass screws for Thermocouple wires plus mini Thermocouple socket -
Easy-to-Read Display
Turn on the backlight to read in dark areas of the plant -
Simple “No Tool” Battery Changes
Four inexpensive “AA” batteries -
Verify Resistance, Cold Junction Temperature & Output
Secondary Display of Resistance for RTDs and compensated millivolt setting and cold junction temperature for T/C.